Wind River Support Network

Meet the Support Network

Home CVE Database CVE-2017-9657

CVE-2017-9657

Description

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Priority: LOW
CVSS v3: 6.5
Component: It doesn't impact WRLinux.
Publish Date: Apr 30, 2018
Related ID: --
CVSS v2: 3.3
Modified Date: Apr 30, 2018

Find out more about CVE-2017-9657 from the MITRE-CVE dictionary and NIST NVD


Products Affected

Login may be required to access defects or downloads.

Product Name Status Defect Fixed Downloads
Linux
Wind River Linux LTS 17 Not Vulnerable -- -- --
Wind River Linux 8 Not Vulnerable -- -- --
Wind River Linux 9 Not Vulnerable -- -- --
Wind River Linux 7 Not Vulnerable -- -- --
Wind River Linux LTS 21 Not Vulnerable -- -- --
Wind River Linux LTS 22 Not Vulnerable -- -- --
Wind River Linux LTS 18 Not Vulnerable -- -- --
Wind River Linux LTS 19 Not Vulnerable -- -- --
Wind River Linux CD release Not Vulnerable -- -- --
Wind River Linux 6 Not Vulnerable -- -- --
Wind River Linux LTS 23 Not Vulnerable -- -- --
VxWorks
VxWorks 7 Not Vulnerable -- -- --
VxWorks 6.9 Not Vulnerable -- -- --
Helix Virtualization Platform Cert Edition
Helix Virtualization Platform Cert Edition Not Vulnerable -- -- --

Related Products

Product Name Status Defect Fixed Downloads

Notes
Requires LTSS - customers must have active LTSS (Long Term Security Shield) Support to receive up-to-date information about vulnerabilities that may affect legacy software. Please contact your Wind River account team or see https://docs.windriver.com/bundle/Support_and_Maintenance_Supplemental_Terms_and_Conditions and https://support2.windriver.com/index.php?page=plc for more information.