Description

An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.

Priority: LOW
CVSS v3: 5.3
Component: It doesn't impact WRLinux.
Publish Date: Feb 13, 2017
Related ID: --
CVSS v2: 2.1
Modified Date: Feb 14, 2017

Find out more about CVE-2016-9355 from the MITRE-CVE dictionary and NIST NVD


Products Affected

Login may be required to access defects or downloads.

Product Name Status Defect Fixed Downloads
Linux
Wind River Linux LTS 17 Not Vulnerable -- -- --
Wind River Linux 8 Not Vulnerable -- -- --
Wind River Linux 9 Not Vulnerable -- -- --
Wind River Linux 7 Not Vulnerable -- -- --
Wind River Linux LTS 21 Not Vulnerable -- -- --
Wind River Linux LTS 22 Not Vulnerable -- -- --
Wind River Linux LTS 18 Not Vulnerable -- -- --
Wind River Linux LTS 19 Not Vulnerable -- -- --
Wind River Linux CD release Not Vulnerable -- -- --
Wind River Linux 6 Not Vulnerable -- -- --
Wind River Linux LTS 23 Not Vulnerable -- -- --
VxWorks
VxWorks 7 Not Vulnerable -- -- --
VxWorks 6.9 Not Vulnerable -- -- --
Helix Virtualization Platform Cert Edition
Helix Virtualization Platform Cert Edition Not Vulnerable -- -- --

Related Products

Product Name Status Defect Fixed Downloads

Notes
Requires LTSS - customers must have active LTSS (Long Term Security Shield) Support to receive up-to-date information about vulnerabilities that may affect legacy software. Please contact your Wind River account team or see https://docs.windriver.com/bundle/Support_and_Maintenance_Supplemental_Terms_and_Conditions and https://support2.windriver.com/index.php?page=plc for more information.
Live chat
Online